Engineering RetiVue’s Adult Model Eye

A photorealistic training model developed through optical design, materials testing and practical prototyping.

CAD cutaway of a black model-eye housing with transparent front optics and a curved retinal image inside.
SolidWorks CAD section of the model eye, showing the housing, optical elements and curved retinal insert.
Client
RetiVue
Project
Adult Model Eye
Biangle Labs contribution
Mechanical engineering, materials and machining expertise, prototyping and iterative development

From the outside, RetiVue’s Adult Model Eye is an unassuming black-and-white assembly. Its purpose becomes clear when you look through it with an instrument used to examine the retina. Instead of painted lines standing in for blood vessels, you see a detailed, photorealistic retinal image through optics designed to reproduce the relevant dimensions and behavior of a human eye.

Creating that view took a three-year collaborative development effort. Optical design was only the beginning. A curved retinal print, a stable optical fluid, suitable plastics, reliable seals and a repeatable manufacturing process all had to work together. Biangle Labs contributed mechanical engineering, materials and machining expertise, prototyping and iterative development alongside RetiVue’s optical, clinical and engineering team.

This is the story of the Adult Model Eye: the problems that appeared between the first design and a useful training product, and the decisions that kept the project moving.

Why training needs a realistic eye

Imagine that you are listening to this article because you can no longer see the text clearly. For many people, eyeglasses or treatment for cataracts can improve vision. The camera analogy is useful: changing a lens can correct some problems, but it cannot repair damaged film. The retina is the light-sensitive tissue at the back of the eye, and problems there can require a very different approach.

Retinal screening gives an eye-care professional a view of that tissue. A technician captures images, and a clinician reviews them for abnormalities. The workflow is familiar from other forms of medical imaging: acquiring a useful image and interpreting it are different tasks, and both matter.

Widefield retinal cameras extend how much of the retina can be examined in the resulting view. RetiVue’s supplied comparison illustrates its EntireEye camera’s wider view beside a narrower field. More area is useful context for the clinician, but obtaining that view requires skill. Camera alignment, pupil size, focus and other factors influence the image that is finally available for review.

Wide retinal image with a smaller field-of-view circle, beside a RetiVue EntireEye camera.
RetiVue’s supplied comparison of widefield and narrower retinal views, beside its EntireEye camera. Open image

Dr. Paul Yates, RetiVue’s founder and an ophthalmic surgeon at the University of Virginia, recognized that watching a training video was no substitute for practicing on the actual equipment. The sensible place to learn its alignment and handling is before a patient is in the chair.

That creates a particular requirement for the training aid. It must do more than resemble an eye on a classroom shelf. It needs to present the optical conditions and retinal detail that the operator will encounter through the imaging instrument. A visually attractive model with the wrong optics would teach the wrong task.

Designing the view through the instrument

RetiVue explored existing model eyes and found that they did not provide the combination of optical accuracy and retinal realism the team needed. Some represented blood vessels with painted lines or sewing thread. Those features can explain anatomy, but they do not produce the detailed retinal view needed for this kind of imaging practice.

The Adult Model Eye was developed around the structures relevant to that view. The cornea and lens needed appropriate shapes, dimensions and optical power. The distances between the cornea, lens and retina also mattered. A transparent filling fluid needed to provide the intended refractive behavior inside the assembly.

Realism included reflections and haze. The goal was not simply to make the clearest possible target for a camera. The materials and optical surfaces needed to reproduce features an operator would see when examining an eye. That required careful choices about what to simulate and what could remain a simplified mechanical structure.

Side-by-side view of a painted eye model with simple vessels and a RetiVue model with detailed retinal vessels.
A painted educational model beside the photorealistic retinal detail used in RetiVue’s model eye. Open image

The photorealistic retinal image completed that combination. Viewed through an indirect ophthalmoscope or an appropriate retinal imaging instrument, it gives an operator meaningful detail to find, align and bring into focus. The model can then be used repeatedly for practice without asking a patient to participate in each learning attempt.

Every design choice returned to the same question: does this help reproduce the view and handling needed for the intended training task? That question linked the optical design to the mechanical work, and it became especially useful when the project moved into less predictable materials problems.

Early optical simulation showing rays through the model-eye optics and curved retina, with a speculum indicated.
Early optical simulation used to explore the model-eye geometry before prototyping. This is the supplied design-stage view. Open image

Making a flat print become a curved retina

One of the hardest challenges was producing a hemispherical retinal shell with a detailed image on its inner surface. The obvious first answer was to print a retinal photograph and attach it to the back of the model. We tried versions of that idea. It did not solve the problem.

You can see why by trying to fit a flat sheet of paper smoothly inside a soup bowl. Wrinkles appear because the sheet cannot conform to the bowl’s curvature without stretching, shrinking or being cut. The model-eye shell had the same geometric difficulty, with a much smaller and more demanding surface.

The team explored transfer decals, printing methods used on objects such as cups and cans, and photolithography. These approaches did not give us a satisfactory way to put the required detail onto this highly curved surface. Getting a recognizable image was not enough; the process had to produce an insert that could perform its optical role inside the model.

The inspiration eventually came from a childhood craft: Shrinky Dinks. A flat plastic sheet can carry a drawing and then change shape when heated. Shrinking also brings the printed dots closer together, increasing their density. That suggested a route from a larger, printable image to the small curved retinal surface we needed.

Shrinky Dinks craft-kit packaging illustrating plastic pieces that shrink and can be shaped when heated.
Shrink-plastic toys helped inspire a process for printing and forming the curved retinal insert. Open image

The resulting approach was to print the retinal image on thermoformable polystyrene and form the material into the required hemisphere. The idea is easy to describe. Making it reliable took about a year of development.

The printer, ink, polystyrene brand and thickness, forming equipment, temperatures and timing all affected the result. The team worked through those variables to establish a repeatable method. The task was not finished when one attractive specimen came out of the process; the method needed to keep producing usable retinal shells.

This approach also made the retinal content adaptable. Prints could depict different retinal disease presentations rather than relying on one permanently painted pattern. The printed detail and the physical geometry could be developed together, giving the model a practical role in both imaging work and training.

Hand-held circular model-eye view showing fine retinal vessels and two bright reflected spots.
Photorealistic retinal detail inside the RetiVue model eye, viewed through an indirect ophthalmoscope. Open image

The materials problems that appeared later

Choosing the outer shell sounds less complicated than creating a photorealistic retina. In practice, it brought its own set of questions. The team tested plastics for machinability, cost, color, thickness, longevity and opacity. The shell had to support the assembly, retain the filling fluid and control unwanted light paths.

A material that was convenient to machine could still be unsuitable after assembly. A fluid that looked clear when first filled could still behave poorly over time. Those decisions could not be made independently: the fluid, printed insert, shell, coatings, lenses and seals formed one system.

In early prototypes, fogging developed behind the iris, bubbles appeared, and the retinal print could be damaged. Some of these failures were not obvious during the first examination. They emerged in models made months earlier, turning an initially useful optical assembly into something that no longer met its purpose.

Over a year went into testing clear substances and filling methods. The requirements included thermal stability, optical suitability, compatibility with the surrounding materials, and avoiding corrosion, toxicity, bubbles and unwanted fogging. Finding a suitable fluid introduced further manufacturing challenges, but it was necessary to preserve the intended reflections, haze and magnification.

The interfaces deserved as much attention as the bulk materials. At the edge of the model cornea, surface roughness had to balance two needs: it needed to be low enough to avoid encouraging cracks, yet sufficient for adhesion and a reliable fluid seal. A small detail at an assembly boundary could decide whether the whole model remained usable.

Close-up of the model’s transparent cornea lens with an inset magnifying its textured edge.
Cornea-lens edge detail. Surface roughness had to balance crack prevention and adhesion for a reliable fluid seal. Open image

The project began with optical geometry and developed into an extended investigation of materials and manufacturing processes. The curved retinal print was a known challenge from the start. The long-term fogging and fluid-loss problems were discovered through building, waiting, inspecting and testing.

That distinction matters. An initial optical simulation can guide a design, but it cannot answer every question about what a filled assembly will do months later. We addressed these later failures because they interfered with the core requirement: a realistic retinal view through the examination instrument.

Cause and effect kept the work focused. If a change affected that view, it deserved attention. If it did not materially affect the intended use, it could be evaluated against the other requirements or recorded for a later version. The question was always what the user needed the model to do, rather than how many features we could fit into it.

A related application in infant training

The retinal-printing and forming approach was also adapted to smaller eyes for a premature-infant training model. This was a related extension of the work, with its own dimensions and use context; it should not be confused with the Adult Model Eye.

Placed in a baby training mannequin, the infant model lets practitioners practice using portable widefield imaging equipment before examining fragile newborns. It illustrates why realistic practice matters when there is little room to learn equipment handling during a clinical examination.

Baby training mannequin with model eyes, wearing a striped sweater.
Related application: the retinal-insert approach was adapted for an infant model used in a baby training simulator. Open image

Turning different kinds of expertise into one product

No single discipline could have delivered the model alone. Optical engineering established the geometry and behavior to reproduce. Clinical expertise explained what mattered during examination and training. Mechanical engineering, materials work, machining and prototyping turned those requirements into an assembly that could be built and used.

Biangle Labs worked on that mechanical and manufacturing side in collaboration with RetiVue. Simplifying the design for manufacture did not mean making the task simple. Machining tolerances, assembly choices, material compatibility and the behavior of the filled model all required iteration.

RetiVue’s wider engineering team and its electrical and software development work provided the imaging-system context. The clinical expert, optical engineer and mechanical team needed a shared understanding of the intended use. Without that exchange, it would have been easy to optimize an individual component while missing the training problem the complete product was supposed to solve.

Collaboration

Different expertise, shared requirements

  1. Clinical subject matter expert

    Defines the use case and the view needed during examination and training.

  2. Biangle Labs mechanical engineering

    Develops the assembly through materials work, machining and prototyping.

  3. RetiVue electrical and software engineering

    Provides the imaging-system context alongside the wider RetiVue team.

Feedback returns to clinical review and the next design decision.

Clinical expertise, RetiVue’s electrical and software work, and Biangle Labs’ mechanical engineering supported the model-eye development. The article also credits the optical engineering that established the model’s geometry.

Three lessons for product development teams

The particular materials and optical challenges make this project distinctive. The decisions behind them also apply to other products built around an unmet need. Three lessons stand out.

1. Give expert input a clear framework

A subject matter expert does more than answer questions. They can reveal requirements the engineering team did not know to ask about. In this project, clinical insight was essential to distinguishing a plausible-looking model from a useful imaging and training tool.

That expertise works best within a defined use case, target market and list of essential features. Without that framework, a discussion can drift toward interesting possibilities that do not serve the current product. The team then spends effort exploring new directions without a clear way to decide whether they belong in the release being built.

This does not mean freezing the scope regardless of what is learned. It means recording new ideas in a backlog and deciding when to act on them. Complete the intended iteration, evaluate what was built, and compare those ideas with the product in hand.

Ask whether a proposed feature offers a compelling advantage. Is it essential to the intended use, or desirable for another use case? Would the product meet its requirements without it? If a backlog item must be incorporated, begin the next iteration with a revised, explicit set of goals rather than letting the change enter unnoticed halfway through the work.

Each cycle needs a physical result whenever the task permits it. Build the prototype. Hold it in your hands. Let other people interact with it and observe what happens. Weeks or months of design work should lead to something that can teach the team more than another discussion or another restart.

One request in this project was a deformable cornea instead of a rigid acrylic one. A softer surface could provide more realistic feedback when practicing with instruments that touch the cornea. It also introduced difficult questions for a fluid-filled model.

The team investigated that possibility, but it was not an essential requirement for the initial use cases, including the related infant model used with contact cameras. The decision was to keep the rigid cornea in the initial product and retain deformability for a future iteration. The idea was useful; its timing was the decision.

Iteration

Finish a cycle. Learn from what you build.

  1. Define goals

    Agree on the use case, target market and essential features.

  2. Gather expert input

    Apply specialist knowledge within that framework.

  3. Develop and build

    Deliver the design set for this iteration.

  4. Evaluate and test

    Hold the prototype, gather feedback and review the requirements.

Use the evidence to define the next iteration.

Define the use case, gather expert input, build and test, and keep future ideas in a backlog.

2. Look for opportunities around the original need

RetiVue’s core business was developing widefield retinal imaging devices. The model eyes first answered an internal research and development need because the team could not find a suitable product. The same model then became a way to train customers on RetiVue’s equipment.

That experience revealed a broader opportunity. The useful feature was not a dependency on one branded camera. It was a realistic eye that could support practice with different retinal examination and imaging tools, from relatively simple instruments to complex camera systems.

Clinics do not replace imaging equipment frequently. Technicians may move between practices and need to learn the equipment available at each one. A training aid with value beyond a single system addresses a practical need without requiring the company to abandon its core technology.

The sequence is worth noticing: an internal problem led to a development tool; the tool supported existing customers; and that experience suggested another market. The broader opportunity grew from the work already done and from understanding how people would use it.

For another product team, the lesson is to look beyond the first application while keeping its requirements clear. There may be adjacent users or training needs that benefit from the same underlying work. Explore them deliberately rather than assuming every possible use must be built into the first version.

Market opportunity

A useful tool can grow beyond its first purpose

  1. Widefield imaging expertise

    RetiVue’s core technology and understanding of retinal imaging.

  2. Internal development need

    A realistic model for research and development.

  3. Customer training

    Practice with the imaging equipment before patient use.

  4. Broader training use

    Potential value across different retinal examination and imaging tools.

An internal development tool became a training product with value beyond one imaging system.

3. Judge completion against requirements

“Good enough” is not simply the point where the team feels satisfied. It is the point where the product meets the requirements set for its intended use. Sometimes reaching that point means discarding months of work or returning to an older design. Sometimes it means resisting another attractive feature because the essential work is already done.

The model-eye shell provides a concrete example. Its initial plastic was selected for machinability, weight, cost and opacity. Months later, the team discovered that the fluid could slowly escape through the material’s porosity. Fluid loss introduced bubbles, which interfered with the view of the retina.

This was a requirement-blocking problem. The team found an insoluble coating for the internal plastic surface that prevented the fluid from escaping. It had to be solved for the model to remain useful; it could not reasonably be treated as an optional refinement.

Compare that with the deformable cornea. Customers had expressed interest in it, but the initial rigid design still met the essential use requirements. The team did not discard that design or delay the first commercial product to add a feature that could be addressed later.

When the only focus group is the small team working on the product, it is easy to become too close to the design. Requirements, deadlines and budgets can start to shift, and the original unmet need becomes less visible. Clear goals and user feedback give the team a way to distinguish an essential correction from another round of development for its own sake.

Persistence therefore includes both the willingness to fix a difficult problem and the discipline to finish. Plan to iterate, define the requirements at the start of each cycle, and deliver something that can be evaluated. A functional product meeting its intended requirements creates a better basis for the next version than an indefinite pursuit of every possible improvement.

A training product built through repeated testing

The result was a photorealistic model eye for imaging practice and research, combining optical design with the less visible work of materials selection, machining, printing, forming, filling and sealing. RetiVue offers the Adult Model Eye Kit as a training and evaluation product, and the model-eye approach is described in a published U.S. patent.

The project team also supplied a real retinal image acquired with the EntireEye Portable Camera after the operator had practiced using a model eye. It shows the clinical imaging task that the related training model was designed to support. The product story rests on the engineering work: reproducing the required view, discovering failures over time, correcting them and delivering the intended function.

Real retinal camera image with branching vessels, optic disc and bright reflected spots.
A real retinal image of a premature infant acquired with the EntireEye Portable Camera after model-eye training, as described by the project team. Open image

For us, the lasting lesson is the value of a shared goal. Clinical guidance, optical design and mechanical execution worked together because the team kept returning to what the user needed to see and do. That gave each experiment a purpose and each design decision a basis.

What does your prototype need to prove?

Build your project brief

A difficult product often starts with a familiar question: can we make this work reliably under the conditions that matter to the user? The answer may require a new process, a material investigation or a carefully scoped prototype before a detailed design is ready.

If you are facing a similar development problem, build a project brief with our quiz. It will help you describe the current stage, important unknowns and the next useful milestone. You can also contact Biangle Labs or explore more of our work.